Health Products

Alinity C system Optics Assembly (2020-05-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.Alinity C system Optics Assembly
Manufacturer
Abbott GmbH
Max Planck Ring 2
Wiesbaden
65205
GERMANY
03R67-01

Introduction

Affected Products

Alinity C system Optics Assembly

Reason

This Product Correction letter is to inform you of an issue with the Alinity c system Optics Assembly (part number C-35016278-01).  Abbott has identified the potential for a nonlinear optics response due to an issue with the Alinity c optics.  The issue is with the diffraction grating within the optics assembly and can result in a lower than expected absorbance value.