RayStation (2020-05-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
4.5.1.14
Manufacturer
RaySearch Laboratories AB (PUBL)
Sveavagen 25
Stockholm
11134
SWEDEN
RayStationNot applicable.

Introduction

Affected Products

RayStation

Reason

This notice concerns an issue found with the proton Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms in RayStation 4, 4.5, 4.7, 5, 6, 7, 8A, 8B, 9A and 9B (including all service packs). The effect of the issue is that robustly optimized plans using tangential fields may exhibit hot spots in the vicinity of the patient surface/skin in the delivered dose. These hot spots would not be visible in dose calculated with the RayStation proton pencil beam dose engine.

The issue can also be relevant for robustness analysis, and dose tracking of non-robust optimized plans using tangential fields. This issue is relevant for any dose computed with the analytical pencil beam (PB) dose engine, and the non-clinical optimization dose of the Monte Carlo (MC) dose engine. The issue is not relevant for the clinical MC dose engine, which handles the tangential fields correctly.