DxA 5000 (2020-06-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Serial # T17000019
Manufacturer
BECKMAN COULTER BIOMEDICAL GMBH
RUPERT-MAYER STREET 44
MUENCHEN,
81379
GERMANY
DxA 5000B50516

Introduction

Affected Products

DxA 5000

Reason

The DxA 5000 instructions for use may be confusing regarding the re-use of sample IDs. Beckman Coulter has received complaints where the re-use of sample barcodes was ""not being detected by the system"". The issue may occur when the original sample tube has been processed and removed from DxA 5000 and a different tube using the same ID has been loaded onto the instrument. While the DxA 5000 does properly detect when duplicate IDs are concurrently being processed, it does not distinguish between a re-run of the same sample and the re-use of the same sample ID with a new sample.

If a sample ID  is re-used with a new sample, the instrument will assume this is the previously processed sample and may incorrectly identify it as having been centrifuged when in fact it has not. This can lead to the processing of inappropriately prepared samples, merged sample history and/or merged sample tracking.