Vitros Chemistry Products System (2020-07-20) Jul 20, 2020 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number All lots. Manufacturer Ortho-Clinical Diagnostics Inc. 100 Indigo Creek Drive Rochester 14626 New York UNITED STATES B. Vitros Chemistry Products System-TBIL Slides (Total Bilirubin) 815 9931 All lots. Manufacturer Ortho-Clinical Diagnostics Inc. 100 Indigo Creek Drive Rochester 14626 New York UNITED STATES C. Vitros XT Chemistry Products-TBIL ALKP Slides 684 4296 All lots. Manufacturer Ortho-Clinical Diagnostics Inc. 100 Indigo Creek Drive Rochester 14626 New York UNITED STATES A. Vitros Chemistry Products System-BUBC Slides (Bilibrubin) 161 2365
Affected Products
Vitros Chemistry Products System-BUBC Slides (Bilibrubin)
Vitros Chemistry Products System-TBIL Slides (Total Bilirubin)
Vitros XT Chemistry Products-TBIL ALKP Slides
There is potential for eltrombopag to interfere with specific Vitros Chemistry product slide assays (listed above). Eltrombopag causes a spectral interferent at the wavelengths that the Vitros TBIL, BUBC and XT TBIL-ALKP slide assays are measured. Eltrombopag was not previously identified as a potential interferent for these Vitros assays.