Sokinox No Delivering and Monitoring System (2020-07-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Article No. 66 94 550
Manufacturer
Maquet Critical Care AB
Rontgenvagen 2
Solna
17154
SWEDEN
Sokinox No Delivering and Monitoring System6694550

Introduction

Affected Products
Sokinox No Delivering and Monitoring System

Reason

Maquet have received reports of corroding pressure sensors resulting in purge failure. If the 'purge failure' alarm occurs during pre-use check (PUC) and PUC cannot be passed the main system should not be used. However, if treatment is needed anyway, the backup system should be used and prior to start of treatment a manual purge procedure must be performed.
To resolve this situation of corroding pressure sensors a correction, i.e. anti-corrosive coating of pressure sensors, has been implemented for circuit board PC2013.