Health Products

Bronchovideoscope (2020-09-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Olympus Medical Systems Corp.
2951 Ishikawa-cho
Hachioji-shi, Tokyo
192-8507
JAPAN
BronchovideoscopeBF-Q180

Introduction

Affected Products

Bronchovideoscope

Reason

Olympus Medical Systems Corporation (OMSC) performed a postmarket risk assessment of the BF-Q180, including adverse events review, which showed that the BF-Q180 is associated with a higher rate of patient infections than other comparable OMSC Bronchoscopes. While this rate of infection is low (less than 0.01%), and patient infection rates depend on a number of factors, out of an abundance of caution and to minimize to the fullest extent possible the risk of infection, OMSC has begun a worldwide transition of the BF-Q180 to newer, similarly indicated Bronchoscope Models.