AMS 700 MS (Momentary Squeeze) Pump (2020-09-24) Sep 24, 2020 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer. Manufacturer American Medical Systems Inc 10700 Bren Road West Minnetonka 55343 UNITED STATES H. CXR PRECONNECT MS 16CM IP IZ 72404266 More than 10 numbers, contact manufacturer. Manufacturer American Medical Systems Inc 10700 Bren Road West Minnetonka 55343 UNITED STATES F. CX PRECONNECT MS 15CM IP IZ 72404235 More than 10 numbers, contact manufacturer. Manufacturer American Medical Systems Inc 10700 Bren Road West Minnetonka 55343 UNITED STATES G. CXR PRECONNECT MS 14CM PS IZ 72404261 More than 10 numbers, contact manufacturer. Manufacturer American Medical Systems Inc 10700 Bren Road West Minnetonka 55343 UNITED STATES D. LGX PRECONNECT MS 12CM IP IZ 72404255 More than 10 numbers, contact manufacturer. Manufacturer American Medical Systems Inc 10700 Bren Road West Minnetonka 55343 UNITED STATES E. CX PRECONNECT MS 18CM PS IZ 72404230 More than 10 numbers, contact manufacturer. Manufacturer American Medical Systems Inc 10700 Bren Road West Minnetonka 55343 UNITED STATES B. PUMP MS IZ 72404209 More than 10 numbers, contact manufacturer. Manufacturer American Medical Systems Inc 10700 Bren Road West Minnetonka 55343 UNITED STATES C. LGX PRECONNECT MS 15CM PS IZ 72404250 More than 10 numbers, contact manufacturer. Manufacturer American Medical Systems Inc 10700 Bren Road West Minnetonka 55343 UNITED STATES A. LGX PRECONNECT MS 18CM PS SKU 72404301
Affected Products
A. LGX PRECONNECT MS 18CM PS SKU
B. PUMP MS IZ
C. LGX PRECONNECT MS 15CM PS IZ
D. LGX PRECONNECT MS 12CM IP IZ
E. CX PRECONNECT MS 18CM PS IZ
F. CX PRECONNECT MS 15CM IP IZ
G. CXR PRECONNECT MS 14CM PS IZ
H. CXR PRECONNECT MS 16CM IP IZ
Product removal of unused inventory of the AMS 700 MS (Momentary Squeeze) Pump following an increase in complaints related to the initial activation of the device. An "initial activation" complaint refers to a problem encountered during the initial activation of the device, typically within the first 2 months after an AMS 700 implant procedure, in which the patient and/or physician is unable or has difficulty activating the pump in order to achieve cylinder inflation, even after exhaustive troubleshooting and patient training.