Siemens Healthcare Diagnostics has identified an issue with the Atellica Solution products (Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer) through investigation of customer complaints for software (SW) versions V1.23.1 (SMN 11485021) or lower.
When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values.