Abbott has identified the ARCHITECT EBV VCA IgM Calibrator lot 08159BE00 may show reduced RLU (Relative Light Units) signal. This reduced RLU signal may cause a decrease in cutoff RLU values in assay calibration resulting in:
upward shift in quality control S/CO values
non-reactive to false grayzone patient results in the range of 0.50 - 0.56 S/CO
grayzone to false reactive patient results in the range of 1.00 - 1.13 S/CO
The effect of decreased cutoff RLU values is observed when using the ARCHITECT EBV VCA IgM Calibrator lot 08159BE00 and is independent from the ARCHITECT EBV VCA IgM Reagent lot used.