ARCHITECT I SYSTEM - EBV VCA IgM Calibrator (2020-11-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
08159BE00
Manufacturer
Abbott GmbH
Max Planck Ring 2
Wiesbaden
65205
GERMANY
ARCHITECT I SYSTEM -  EBV VCA IgM Calibrator3P66-01

Introduction

Affected Products

ARCHITECT I SYSTEM -  EBV VCA IgM Calibrator

Reason

Abbott has identified the ARCHITECT EBV VCA IgM Calibrator lot 08159BE00 may show reduced RLU (Relative Light Units) signal. This reduced RLU signal may cause a decrease in cutoff RLU values in assay calibration resulting in:
  • upward shift in quality control S/CO values
  • non-reactive to false grayzone patient results in the range of 0.50 - 0.56 S/CO
  • grayzone to false reactive patient results in the range of 1.00 - 1.13 S/CO


The effect of decreased cutoff RLU values is observed when using the ARCHITECT EBV VCA IgM Calibrator lot 08159BE00 and is independent from the ARCHITECT EBV VCA IgM Reagent lot used.