Health Products

Regional Block Needles 22G x 1.50in (2020-12-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0127675
Manufacturer
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
UNITED STATES
Regional Block Needles 22G x 1.50in408348

Introduction

Affected Products

Regional Block Needles 22G x 1.50in

Reason

Regional anesthesia needles instructions for use (IFU) (document# 10000354809 rev. 01) currently lists the following contraindication which was not approved nor reported to notified body (nsai) prior implementation:

"the risks associated with the use of this device may be increased due to patient physiological characteristics and clinical needs".

The initial contraindication(s) listed below were deleted/changed.

"These needles should not be used if there is:
  • infection at the intended site of entrance
  • allergy to the intended anesthetic agent
  • evidence of a severe coagulation disorder
  • sepsis
  • abnormal anatomy that would preclude successful placement of the needle
  • the patient is uncooperative or unable to be positioned adequately"