Health Products

NM (2020-12-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
830Y60095B. NM 830
Manufacturer
GE Medical Systems Israel,
Functional Imaging
4 Hayozma Street
Tirat Hacarmel
0030200
ISRAEL
H3910AB
870Y64085A. NM/CT 870 DR
Manufacturer
GE Medical Systems Israel
Functional Imaging
4 Hayozma Street
H3100AR

Introduction

Affected Products 

A. NM/CT 870 DR

B. NM 830

Reason

Rotor bearing screws were found loose on one detector, leading to the release of one of the four rotor bearings in the detector. In the unlikely event that the screws loosen, and multiple other design redundancies also fail, the detector could fall and lead to potentially life-threatening bodily harm. 
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime