Health Products

Phadia System - ELIA DSDNA (2020-12-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Department: Health Canada
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
BFA3Y/0142
Manufacturer
PHADIA AB
RAPSGATAN 7, P.O. BOX 6460
UPPSALA
75137
SWEDEN
B. Phadia 250 System- ELIA DSDNA14-5500-01
BFA3Y/0142
Manufacturer
PHADIA AB
RAPSGATAN 7, P.O. BOX 6460
UPPSALA
75137
SWEDEN
A. PHADIA 100 SYSTEM- ELIA DSDNA14-5500-01

Introduction

Affected Products 

A. PHADIA 100 SYSTEM- ELIA DSDNA

B. Phadia 250 System- ELIA DSDNA

Reason

Customer complaints have reported a decrease of values for EliA ANA positive control when using EliA dsDNA well lot bfa3y/0142. An internal investigation showed that EliA dsDNA well lot bfa3y/0142 gives higher results, and in some cases the results may be 3 times higher than other dsDNA well lots. There is potential for a false positive test result.