Health Products

Penumbra JET 7 Reperfusion Catheter and Penumbra Hi-Flow Aspiration Tubing (2020-12-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Penumbra Inc.
1351 Harbor Bay Parkway
Alameda
94502
California
UNITED STATES
B. Penumbra JET 7 Reperfusion Catheter and Penumbra Hi-Flow Aspiration Tubing5MAXJET7KIT
All lots.
Manufacturer
Penumbra Inc.
1351 Harbor Bay Parkway
Alameda
94502
California
UNITED STATES
A. Penumbra JET 7 Reperfusion Catheter 5MAXJET7

Introduction

Affected Products
  1. Penumbra JET 7 Reperfusion Catheter
  2. Penumbra JET 7 Reperfusion Catheter and Penumbra Hi-Flow Aspiration Tubing

Reason

The Catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization or contrast injection may result in potential vessel damage, and subsequent patient injury or death.