Health Products

POLYAXIAL SCREW (2020-12-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
KRBX-302665POLYAXIAL SCREW
Manufacturer
K2M, Inc.
600 Hope Parkway SE
Leesburg
20175
Virginia
UNITED STATES
7601-04026

Introduction

Affected Products 

 

POLYAXIAL SCREW

Reason

A limited number of screws in a single lot (KRBX) of Yukon Polyaxial Screws- catalog number 7601-04026, size 4.0X26mm-were manufactured with 4.0X26mm dimensions and a dark blue shank, but incorrectly labeled (direct marked on screw) as catalog number 7601-05024, size 5.0X24mm. Therefore, these screws are longer and narrower than marked. Stryker has not received any reports of harm to patients associated with this nonconformance.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime