Health Products

Cobas (2020-12-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All software versions. 
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
B. COBAS C 513 ANALYZERNot applicable. 
1924-01
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
C. COBAS PRO SAMPLE SUPPLY UNITNot applicable. 
All software versions. 
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
A. Cobas 8000 Core Unit5641446001

Introduction

Affected Products 
  1. Cobas 8000 Core Unit
  2. COBAS C 513 ANALYZER
  3. COBAS PRO SAMPLE SUPPLY UNIT

Reason

In very rare circumstances, a database timeout error can occur which affects the settings stored on the SQL server leading to a deactivation of some analyzer settings on Cobas 8000, Cobas C 513 and Cobas Pro Systems.
Get told about the next health product recall.
Get health product recall alerts