ORTHO VISION ANALYZER (2020-12-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Software ver 5.12.8 and 5.13.0
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS INC.
1001 US HWY 202
Raritan
08869
New Jersey
UNITED STATES
B. ORTHO VISION MAX ANALYZER6904576
Software ver 5.12.8 and 5.13.0
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS INC.
1001 US HWY 202
Raritan
08869
New Jersey
UNITED STATES
A. ORTHO VISION ANALYZER6904577

Introduction

Affected Products 

A. ORTHO VISION ANALYZER
B. ORTHO VISION MAX ANALYZER

Reason

Due to a software anomaly on Ortho Vision and Ortho Vision Max Analyzers configured with software version 5.12.8 Or 5.13.0, flushing of the pipette may occur in a reagent vial after the Pipette Arm (PIPA) exits idle mode, one or more reagent/diluent vials are loaded on the system and the operator enters maintenance mode before the analyzer completes the reagents inventory.