Health Products

EPIQ ULTRASOUND SYSTEM (2021-01-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots. B. EPIQ 7 ULTRASOUND SYSTEM
Manufacturer
Philips Ultrasound Inc.
22100 Bothell-Everett Highway
Bothell
98021-8431
Washington
UNITED STATES
EPIQ 7
All lots.C. EPIQ DIAGNOSTIC ULTRASOUND SYSTEM
Manufacturer
Philips Ultrasound Inc.
22100 Bothell-Everett Highway
Bothell
98021-8431
Washington
UNITED STATES
EPIQ
All lots. A. EPIQ 5 ULTRASOUND SYSTEM
Manufacturer
Philips Ultrasound Inc.
22100 Bothell-Everett Highway
Bothell
98021-8431
Washington
UNITED STATES
EPIQ 5

Introduction

Affected Products

A. EPIQ 5 ULTRASOUND SYSTEM
B. EPIQ 7 ULTRASOUND SYSTEM
C. EPIQ DIAGNOSTIC ULTRASOUND SYSTEM

Reason

Philips recently discovered an issue associated with EPIC ultrasound systems where the control panel arm assembly could have missing or loose screws. In these cases, if undue force, pressure or weight is applied, the control panel mechanism can fail and break off. We are aware of three instances where the control panel arm with missing/loose screws broke and collapsed, one of them causing a minor injury to the operator.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime