Health Products

Custom Tubing Packs (2021-01-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer. 
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
F. PERFUSION SET WITH BIOLINE COATINGBEQ-HQV 48707
More than 10 numbers, contact manufacturer. 
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
D. CARDIOPLEGIA SET WITH SOFTLINE COATINGBO-H 95260
More than 10 numbers, contact manufacturer. 
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
E. PERFUSION TUBING PACK WITH BIOLINE COATINGBEQ-H 48730
More than 10 numbers, contact manufacturer. 
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
B. CUSTOM PACK WITH SOFTLINEBO-HQV 66604
More than 10 numbers, contact manufacturer. 
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
C. CARDIOPLEGIA SET WITH BIOLINE COATINGBEQ-H 48761
More than 10 numbers, contact manufacturer. 
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
A. HEART LUNG PERFUSION PACK WITH SOFTLINEBO-HQV 48707

Introduction

Affected Products 

A. HEART LUNG PERFUSION PACK WITH SOFTLINE
B. CUSTOM PACK WITH SOFTLINE
C. CARDIOPLEGIA SET WITH BIOLINE COATING
D. CARDIOPLEGIA SET WITH SOFTLINE COATING
E. PERFUSION TUBING PACK WITH BIOLINE COATING
F. PERFUSION SET WITH BIOLINE COATING

Reason

Maquet Cardiopulmonary GmbH became aware of a missing cable tie in the cardioplegia (Plegiox) line to secure the connection between connector and tubing next to the roller pump for the article numbers subject of this FSN.  A missing cable tie could lead to a leakage and/or loosening of the tube in the cardioplegia (Plegiox) line.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime