VITROS Chemistry Products VersaTips (2021-01-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
413420
Manufacturer
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
VITROS Chemistry Products VersaTips10758750006786

Introduction

Affected Products 

VITROS Chemistry Products VersaTips

Reason

Ortho Clinical Diagnostics received customer complaints for the following condition codes on a Vitros 5600 system:
TMN-410: UIA Metering Aspirate Servo - pump travel limit reached
TEN-410: Uslide Metering Aspirate Servo - pump travel limit reached
TML-49A: UIA Metering Cuvetip seal failure at 1 tip sealer
Initial investigation found that the increase in these condition codes was related to improper tip sealing when using one specific lot of VITROS VersaTips. Upon inspection, a deformity was noted in some tips from this lot. The issue is intermittent as not all tips from this lot are affected. No other lots appear to be affected. This deformity can lead to an improper seal of the tip on the proboscis. An improperly sealed tip may cause a condition code to post but will not always post a condition codes or suppress result reporting.