Ortho Clinical Diagnostics received customer complaints for the following condition codes on a Vitros 5600 system:
TMN-410: UIA Metering Aspirate Servo - pump travel limit reached
TEN-410: Uslide Metering Aspirate Servo - pump travel limit reached
TML-49A: UIA Metering Cuvetip seal failure at 1 tip sealer
Initial investigation found that the increase in these condition codes was related to improper tip sealing when using one specific lot of VITROS VersaTips. Upon inspection, a deformity was noted in some tips from this lot. The issue is intermittent as not all tips from this lot are affected. No other lots appear to be affected. This deformity can lead to an improper seal of the tip on the proboscis. An improperly sealed tip may cause a condition code to post but will not always post a condition codes or suppress result reporting.