LifePak CR2 Defibrillator (2021-02-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer. 
Manufacturer
Physio-Control, Inc.
11811 Willows Road North East
Redmond
98052
Washington
UNITED STATES
LifePak CR2 DefibrillatorMore than 10 numbers, contact manufacturer.

Introduction

Affected Products 

LifePak CR2 Defibrillator

Reason

Stryker has become aware that the lid magnet for this device can occasionally become separated and lost. Until recently, Stryker was previously only aware that if the lid magnet is missing, the device would not automatically turn on and off when the lid is opened and closed. Stryker has now become aware that if the lid magnet is missing, it also causes the LIFEPAK CR2 battery to deplete more quickly resulting in premature battery life. (Complaint trending from April 30 2020 to November 1 2020). The LPCR2 Operating Instructions currently provide users with no warning that the battery can become prematurely depleted if the lid magnet is missing.