Health Products

Straumann Jason Membrane (2021-02-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
19KA51600Straumann Jason Membrane
Manufacturer
Institut Straumann AG,
Peter Merian-Weg 12,
Basel
4002
SWITZERLAND
AW-681520

Introduction

Affected Products 

Straumann Jason Membrane

Reason

The listed batches of Jason Membrane distributed to and in Canada, were subjected to a raw material supplier change which has not been sufficiently evaluated. The supplier has introduced a change on the animal tissue sourcing without appropriate notification. After becoming aware of this change, it was determined that it had been implemented without the appropriate amendment notification to Health Canada License No. 101739