Health Products

Implantable Cardioverter Defibrillator (2021-02-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
H. EVERA S VR DIGITAL SINGLE CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATORDVBC3D1
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
I. BRAVA QUAD CRT-DDTBC2Q1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
F. EVERA S DR DIGITAL DUAL CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATORDDBC3D1
BWU600507S
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
O. VISIA AF S VRDVAC3D1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
P. VISIA AF MRI XT VR SURESCANDVFB2D4
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
G. EVERA XT VR DIGITAL SINGLE CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATORDVBB2D1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
Q. VISIA AF MRI S VR SURESCANDVFC3D4
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
D. VIVA QUAD XT CRT-DDTBA2Q1
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
E. EVERA XT DR DIGITAL DUAL CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATORDDBB2D1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
R. AMPLIA MRI QUAD CRT-D SURESCANDTMB2QQ
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
B. VIVA S CRT-D DIGITAL IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WITH CARDIAC RESYNCHRONIZATION THERAPYDTBB2D1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
K. EVERA MRI XT DR SURESCANDDMB2D1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
L. EVERA MRI XT VR SURESCANDVMB2D1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
C. BRAVA CRT-D DIGITAL IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WITH CARDIAC RESYNCHRONIZATION THERAPYDTBC2D1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
M. EVERA MRI S DR SURESCANDDMC3D1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
N. EVERA MRI S VR SURESCANDVMC3D1
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
A. VIVA XT CRT-D DIGITAL IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WITH CARDIAC RESYNCHRONIZATION THERAPYDTBA2D1
More than 10 numbers, contact manufacturer. 
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
J. VIVA QUAD S CRT-DDTBB2QQ

Introduction

Affected Products 
  1. VIVA XT CRT-D DIGITAL IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WITH CARDIAC RESYNCHRONIZATION THERAPY
  2. VIVA S CRT-D DIGITAL IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WITH CARDIAC RESYNCHRONIZATION THERAPY
  3. BRAVA CRT-D DIGITAL IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WITH CARDIAC RESYNCHRONIZATION THERAPY
  4. VIVA QUAD XT CRT-D
  5. EVERA XT DR DIGITAL DUAL CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
  6. EVERA S DR DIGITAL DUAL CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
  7. EVERA XT VR DIGITAL SINGLE CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
  8. EVERA S VR DIGITAL SINGLE CHAMBER IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
  9. BRAVA QUAD CRT-D
  10. VIVA QUAD S CRT-D
  11. EVERA MRI XT DR SURESCAN
  12. EVERA MRI XT VR SURESCAN
  13. EVERA MRI S DR SURESCAN
  14. EVERA MRI S VR SURESCAN
  15. VISIA AF S VR
  16. VISIA AF MRI XT VR SURESCAN
  17. VISIA AF MRI S VR SURESCAN
  18. AMPLIA MRI QUAD CRT-D SURESCAN

Reason

Notification informing customers of a potential issue for a subset of Implantable Cardioverter Defibrillators (ICDs) and Cardiac Esynchronization Therapy Defibrillators (CRT-DS). Medtronic has identified that a small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation. The subset of ICDs and CRT-DS affected by this issue were last implanted in February 2019 and manufactured with a specific battery design that is no longer being distributed.

The rapid depletion is caused by a latent shorting mechanism involving lithium plating, resulting from a thermal gradient between the anode and cathode components of the battery. Devices with higher pacing outputs and high pacing percentages (e.g. CRT-D devices) have the lowest probability of occurrence. Conversely, devices with low current drain (evidenced by a longer overall service time from implant to RRT) have a higher probability of experiencing this issue. The probability of this issue developing is constant after approximately three years of service time.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime