ALINITY I HIV AG/AB COMBO CALIBRATOR (2021-02-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
21292BE00ALINITY I HIV AG/AB COMBO CALIBRATOR
Manufacturer
Abbott GmbH
Max Planck Ring 2
Wiesbaden
65205
GERMANY
08P0701

Introduction

Affected Products 

ALINITY I HIV AG/AB COMBO CALIBRATOR

Reason

Abbott identified internally a manufacturing issue where individual vials of the Alinity i HAVAb IgG Calibrator were mislabeled as Alinity i HIV Ag/Ab Combo Calibrator lot number 21292BE00.

Impacted vials can cause invalid calibrations due to low RLU (Relative Light Units) calibrator signals. When calibrating using an impacted vial you will observe either:
  • Message code: 1400: Calibration failed for assay (HIV Ag/Ab) number (396). (CV too large for Cal 1).
  • Or, if a calibration passes, the quality control values will be out of specification high and the calibrations are invalid.