Abbott identified internally a manufacturing issue where individual vials of the Alinity i HAVAb IgG Calibrator were mislabeled as Alinity i HIV Ag/Ab Combo Calibrator lot number 21292BE00.
Impacted vials can cause invalid calibrations due to low RLU (Relative Light Units) calibrator signals. When calibrating using an impacted vial you will observe either:
Message code: 1400: Calibration failed for assay (HIV Ag/Ab) number (396). (CV too large for Cal 1).
Or, if a calibration passes, the quality control values will be out of specification high and the calibrations are invalid.