MAGNUS 1180 OPERATING TABLE COLUMNS (2021-02-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Maquet GmbH
Kehler Strasse 31
Rastatt
76437
GERMANY
MAGNUS 1180 OPERATING TABLE COLUMNS118001

Introduction

Affected Products 

MAGNUS 1180 OPERATING TABLE COLUMNS

Reason

Under certain conditions, they have found that a problem may cause the operating table not to function as intended. The tables may develop a hydraulic leakage at the inclination function. This leakage leads to an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning. This malfunction only exhibits when tilt movement is used. They are aware of instances where this problem led to patients having to be transferred to another OR table which prolonged the procedure.