Affected Products
Bravo reflux capsule and calibration-free reflux capsules
This voluntary recall is being conducted following customer reports of the Bravo Reflux Capsule failing to attach to the esophageal mucosa. In cases where the capsule fails to attach to the esophageal mucosa, the potential exists for aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received in the last 2 years. Manufacturing process improvements have been implemented to address this issue.