Health Products

Leckey MyWay Gait Trainer/Walker (2021-03-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
LEC/262983Leckey MyWay Gait Trainer/Walker (2021-03-08)
Manufacturer
JAMES LECKEY DESIGNS LTD
A BALLINDERRY ROAD,
LISBURN
BT2825A
IRELAND
173-1600

Introduction

Affected Products 

Leckey MyWay Gait Trainer/Walker

Reason

The rear legs of the MyWay walker each have a boss into which castors are screwed. This boss can become loose causing the castor to loosen from the rear leg of the chassis. Investigations have confirmed that this is due to variability in the weld process used to attach the boss to tube.  The manufacturer has identified a potential for patient harm.  In the interest of user safety, a decision has been taken to issue a Field Safety Notice proactively (FSN-01-03-21). This will involve a MyWay Walker Rear Leg replacement to be carried out in the field.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime