Health Products

CUSTOM PACK WITH SOFTLINE (2021-03-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
3000146594CUSTOM PACK WITH SOFTLINE
Manufacturer
MAQUET CARDIOPULMONARY GMBH
KEHLER STR. 31, RASTATT, GERMANY, 76437
RASTATT
76437
GERMANY
BO-HQV 48412

Introduction

Affected Products

CUSTOM PACK WITH SOFTLINE

Reason

The sterile pouch (component #700007026) is used to pack the accessories that are placed inside the primary sterile barrier (blister) and intended to organize the articles inside the primary packaging. The pouch #700007026 is used for packing the centrifugal pump ""rotaflow"" and the manifold line,
which are not pre-connected to the tubing set.
The ability of the sterile pouch to maintain its integrity as a sterile barrier during storage and transport is insufficiently demonstrated. The sterility of the device itself is however not affected as the integrity of the primary sterile barrier (blister) is not compromised by storage and transport conditions.