Health Products

Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing, and LogiCal Kits (2021-03-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
3971467B. STOPCOCKS
Manufacturer
Smiths Medical ASD Inc.
6250 Shier Rings Rd
Dublin, Ohio
43016
Ohio
UNITED STATES
MX1431MRL
4042461C. LOGICAL@3 GANG MANIFOLD 1/EA
Manufacturer
Smiths Medical ASD Inc.
6250 Shier Rings Rd
Dublin, Ohio
43016
Ohio
UNITED STATES
M20178R2
3952480A. HIGH PRESSURE ROTATOR WITH MALE LUER LOCK 50/CA
Manufacturer
Smiths Medical ASD Inc.
6250 Shier Rings Rd
Dublin, Ohio
43016
Ohio
UNITED STATES
MX496HP

Introduction

Affected Products 

A. HIGH PRESSURE ROTATOR WITH MALE LUER LOCK 50/CA
B. STOPCOCKS
C. LOGICAL@3 GANG MANIFOLD 1/EA

Reason

Smiths Medical became awarethatspecific lots of Medex® High PressureAdaptors, Rotators, Manifolds,Stopcocks and Tubing and Medex®LogiCal® Kits may have beenmanufactured with insufficient orincomplete welds. Devices with aninsufficient or incomplete weld maypotentially fail during use.