Philips Achieva & Ingenia MR systems (2021-04-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
F. INGENIA ELITION X781358
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
G. INGENIA AMBITION X781356
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
D. INGENIA 3.0T3.0T
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
E. INGENIA 3.0T CX781271
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
B. ACHIEVA 1.5T MRI SYSTEM-MAIN UNITACHIEVA 1.5T
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
C. INGENIA 1.5T1.5T
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. ACHIEVA 3.0T MRI SYSTEM-MAIN UNITACHIEVA 3.0T

Introduction

Affected Products 

A. ACHIEVA 3.0T MRI SYSTEM-MAIN UNIT
B. ACHIEVA 1.5T MRI SYSTEM-MAIN UNIT
C. INGENIA 1.5T
D. INGENIA 3.0T
E. INGENIA 3.0T CX
F. INGENIA ELITION X
G. INGENIA AMBITION X

Reason

For the following MR systems manual in (or out) movement of the tabletop is not completely possible: Ingenia 1.5T, Ingenia 1.5T (s), Ingenia 1.5T CX, Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Ambition S, Ingenia Ambition X, Ingenia Elition S, Ingenia Elition X, Smartpath to Dstream 1.5T, Smartpath to Dstream for XR and 3.0T.
The horizontal tabletop movement can be either motorized or in manual mode.
Manual mode means that the tabletop is moved manually by pushing or pulling the tabletop handgrip by the operator.
When the horizontal movement is in manual mode, the tabletop outward motion may be blocked while the table is only partly out of the bore. This is due to a mechanical failure (wrong gear profile). An unacceptable hazardous situation could occur, when the tabletop is in 'manual' mode during a quick evacuation of the patient from the MR system.