Panther Fusion SARS-CoV-2 Primer Probe Reagent (2021-04-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
291645, 292447
Manufacturer
Hologic, Inc.
10210 Genetic Drive
San Diego
92121
California
UNITED STATES
Panther Fusion SARS-CoV-2 Primer Probe ReagentPRD-06391

Introduction

Affected Products

Panther Fusion SARS-CoV-2 Primer Probe Reagent

Reason

Between April 12, 2021 and April 14, 2021, Hologic's Canadian importer, Accuristix, accidentally shipped to 6 customers product (2 lots of Panther Fusion SARS-CoV-2 Assay PPR Solution) that had experienced temperature excursions during shipment from Hologic to Accuristix. Risk assessments were performed by Hologic for the lots and it was concluded that they were not fit for use.  Hologic informed Accuristix on April 28, 2021 and that is when it was discovered that Accuristix had accidentally shipped the lots to customers.