Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
VA40D
Manufacturer
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY
SYNGO DYNAMICSVA40

Introduction

Affected Products 

SYNGO DYNAMICS

Reason

Multiple Syngo Dynamics VA40D feature enhancements and the resolution to various software defects provided with this software are listed below:                                                                                                                                                                                                                   • Check conflicts does not work consistently across all the servers
• Data mining tool does not populate the calcs which have more than onecontextand present inmultiple departments
• CVIM # (stress testing) Reporting crashed when continuously slickingthe report toggle button inthumbnail mini toolbar
• RetrieveStudies.pl does not include database name in connectionstrings
• LV vol d, MOD A4Cdoes not have all the required disk measurements andother requiredmeasurements in the library.
• If a measurement namehas a different case with that of DB, SR importstops at thatmeasurement
• VA20C additional fetus isgetting created when SR objects are sentkeeping the worksheet andreport open.
• Studies are disappearing fromdatabase
• VA20F fixup tool still completesfixup and merges study even ifclicking on "X" in upper corneronthe possible merge dialog box.