Health Products

SafeScraper & Micross (2021-05-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
15-17018
Manufacturer
C.G.M. S.P.A. DIVISIONE MEDICALE META
VIA E.VILLA, 7
ITALY
B. SafeScraper Twist Curve3987
24-29118
Manufacturer
C.G.M. S.P.A. DIVISIONE MEDICALE META
VIA E.VILLA, 7
ITALY
A. Micross4049

Introduction

Affected Products 

A. Micross
B. SafeScraper Twist Curve

Reason

C.G.M. Spa Divisione Medicale Meta received reports from Steril Milano informing about deviations of sterilization process parameters conducted by steril milano. Following an internal investigation conducted on the basis of last information received from the contract steriliser, C.G.M. Spa Divisione Medicale Meta has established that the sterility of all products batches cannot be guaranteed. 
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime