Health Products

QUADROX-I NEONATAL OXYGENATOR WITH INTEGRATED HARDSHELL RESERVOIR (2021-05-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer. QUADROX-I NEONATAL OXYGENATOR WITH INTEGRATED HARDSHELL RESERVOIR
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
BO-VKMO 11000

Introduction

Affected Products 

QUADROX-I NEONATAL OXYGENATOR WITH INTEGRATED HARDSHELL RESERVOIR

Reason

During design verification testing of Neonatal VKMO's and VHK's a potential impairment of the sterile package of the accessory part ""tight caps"" was identified. The exposure of either the blood path or sterile (operative) field to nonsterile tight caps may cause none, some, or all of the following immediate and/or long-range health consequences (injuries or illnesses):
• Exposure of the sterile (operative) field to an unsterile product which could compromise the entire surgical field.
• Exposure of a patient's vascular system to an unsterile product resulting in inflammation and/or infection.
• Sepsis due the unmitigated and unknown propagation of an infection to other regions of the corpus.

Maquet Cardiopulmonary has not received any complaints or reports of adverse events due to damage to the sterile barrier system of tight caps.