Health Products

Iron Gen.2 (2021-05-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots. 
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
Iron Gen.203183696122

Introduction

Affected Products

Iron Gen.2

Reason

The manufacturer informed us, following customer complaints regarding the increased recovery of controls and discrepant elevated results for the IRON2 on cobas c 311/501/502 and on COBAS INTEGRA 400 plus, of throughput dependent signal drifts with our assay IRON 2 (cat# 03183696122) on cobas c311, cobas c 501/502 and COBAS INTEGRA®400 plus analyzers.