Health Products

T2 TIBIAL NAIL SYSTEM- NAIL INSERTION SLEEVE ELASTIC (2021-05-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
K03CC4FT2 TIBIAL NAIL SYSTEM- NAIL INSERTION SLEEVE ELASTIC
Manufacturer
Stryker Trauma GMBH
Prof.-Kuntscher-Str.1-5
Schoenkirchen
24232
GERMANY
1806-1406S

Introduction

Affected Products 

T2 TIBIAL NAIL SYSTEM- NAIL INSERTION SLEEVE ELASTIC

Reason

Stryker Trauma GmbH (Trauma & Extremities Division) determined through its quality system (product complaints received December 2020) that two specific lots of T2 Nail Insertion Sleeve instrument, referenced in the table above, contain devices that have a different inner diameter than the diameter specified on the outer box label. Specifically, the products in: (i) lot number K03CC50 contained products with an inner diameter that was 2mm smaller than the label indicates; and (ii) lot number K03CC4F contained products with an inner diameter that was 2mm larger than the label indicates.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime