COBAS 6500 ANALYZER (2021-05-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
software version ? 2.3.6.B. COBAS 6500 - COBAS U701 MICROSCOPY ANALYZER
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
6390501001
software version ? 2.3.6.A. COBAS 6500 U601 URINALYSIS SYSTEM - URINE ANALYZER
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
6390498001

Introduction

Affected Products 

A. COBAS 6500 U601 URINALYSIS SYSTEM - URINE ANALYZER
B. COBAS 6500 - COBAS U701 MICROSCOPY ANALYZER

Reason

Our manufacturer confirmed that, under rare circumstance, possible sample mismatch may occur while using the cobas u 701 microscopy analyzer in combination with cobas u 601 urine analyzer with software version 2.3.6. In case of asymptomatic patients (e.g. microhematuria), medical decisions may be affected if a result is incorrectly positive or negative. Therefore, a medical risk for the population at greatest risk cannot be excluded. This issue impacts only customers using cobas u601 in combination with cobas u701. Standalone instruments (cobas u601 used alone) are not impacted by the issue.