Health Products

AVANTI+ CARDIOLOGY SHEATHS (2021-06-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
17919780
Manufacturer
Cordis Corporation
14201 NW 60TH AVE.
MIAMI LAKES
33014
Florida
UNITED STATES
AVANTI+ CARDIOLOGY SHEATHS504657A

Introduction

Affected Products 

AVANTI+ CARDIOLOGY SHEATHS

Reason

Cordis became aware that the catalog 504-657A, size 7F MS (approximately 7.5F) is mislabeled (both outer and inner labels) as size 7F because of a missing “Mid Size” (“MS”) designation. The product in the mislabeled packaging is larger (both inner and outer diameter) than the labeling indicates. The difference (0.5F) represents approximately 0.006 inches.