Health Products

PATIENT DATA MODULE (2021-06-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer. 
Manufacturer
GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES INC
9900 INNOVATION DRIVE
WAUWATOSA
Wisconsin
UNITED STATES
PATIENT DATA MODULE2042084-001

Introduction

Affected Products 

PATIENT DATA MODULE

Reason

If the PDM with Masimo SET® technology has not been powered down completely in more than 2 years, it can display a frozen inaccurate SpO2 saturation value which no longer changes to reflect the patient’s clinical condition. This issue could result in missed SpO2 alarms and delayed diagnosis and treatment.