BIPAP A30 SYSTEM-VENTILATOR & SYSTEM ONE A-SERIES HEATED HUMIDIFIER
BIPAP A40, CANADA
BIPAP A40, CANADA, CORE PACKAGE
Reason
Philips has become aware of two (2) issues that may pose a risk for patients or users of Philips Respironics branded Continuous Positive Airway Pressure (CPAP), Bi-Level Positive Airway Pressure (BIPAP), and Mechanical Ventilators:
Philips has determined from user reports and testing that the Polyester-Based Polyurethane (PE-PUR) sound abatement foam used in Philips continuous and non-continuous ventilators may degrade under certain circumstances, and the degraded particles could potentially enter the air pathway of the device. This issue affects Philips Respironics branded CPAP’s, Bi-Levels, and Mechanical Ventilators.
The results of testing performed by Philips indicate that the PE-PUR sound abatement foam used in these devices may emit certain chemicals. Our investigation to date indicates that this emission occurs during initial operation and may possibly continue throughout the device's useful life.
These issues impact all device product platforms manufactured with Polyester Polyurethane (PE-PUR) sound abatement foam. There is no specific population of device serial numbers which are impacted.