Health Products

Soltive Premium Superpulsed Laser System (2021-06-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Gyrus Acmi, Inc.
9600 LOUISIANA AVE. NORTH
BROOKLYN PARK
55445
Minnesota
UNITED STATES
Soltive Premium Superpulsed Laser SystemTFL?PLS

Introduction

Affected Products 

Soltive Premium Superpulsed Laser System

Reason

The manufacturer, Gyrus Acmi Inc., received complaints about loss of kidney function after lithotripsy procedures involving the subject device. Through a clinical investigation and consultation with medical professionals, the events were potentially the result of knowledge-based failure where the user accustomed to using a different laser (Holmium:YAG) applied more energy than needed to achieve a certain outcome with the Soltive laser. Prior to these reported events, Olympus had planned for a software upgrade containing a new preset for ureteral stone. In light of this investigation, Olympus will expedite the distribution of the software upgrade as a preventative
measure. Additionally, the instruction for use will be updated to include additional verbiage regarding procedural differences between Holmium:YAG lasers and soltive lasers as it relates to device settings.