Health Products

PEDIASAT OXIMETRY CATHETER SET (2021-07-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
62728569PEDIASAT OXIMETRY CATHETER SET
Manufacturer
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
California
UNITED STATES
XT3515SP

Introduction

Affected Products 

PEDIASAT OXIMETRY CATHETER SET

Reason

Edwards has received three (3) customer complaints reporting leaks during use. One complaint noted that after administering medication, the physician flushed a different lumen than the one where the medication was administered, and the patient’s vital signs responded to the medication. The other two complaints noted external leakage at the transition from the green hub (backform) to the tube. Edwards has confirmed a manufacturing defect leading to possible interlumen leakage.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime