A. TRILOGY EVO
B. TRILOGY EVO, O2
C. TRILOGY EV300
Reason
Correction of the impacted devices is required to address software issues which may result in an increase in the baseline pressure. There are two (2) separate issues which can result in a pressure increase in certain applications of the impacted devices:
Issue 1 - EFS calibration pressure increase
In cases where the infant / pediatric External Flow Sensor (EFS) is used with an active flow or dual limb circuit where a manual circuit calibration is performed, an increase in the baseline pressure can occur.
Issue 2 - Pressure drift (continuous use)
Where target pressure applied to the patient can slowly increase over time with continual uninterrupted use due to a floating-point arithmetic rounding error (~2cmh2o/month). This issue was evaluated separately with consideration to devices intended for home use and those intended for the hospital use.