Health Products

DXA 5000 (2021-07-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
T23000022
Manufacturer
BECKMAN COULTER BIOMEDICAL GMBH
RUPERT-MAYER STREET 44
MUENCHEN,
81379
GERMANY
DXA 5000B57634

Introduction

Affected Products

DXA 5000

Reason

Sample carriers (with or without samples) are getting routed to the outer lane of the Single Drive Transport Segment (B57634) to the end of the lane, then dropping off the lane.
Get told about the next health product recall.
Get health product recall alerts