The manufacturer, Eco-Med Pharmaceuticals, Inc., has initiated a recall and quarantine of certain lots of EcoGel 200 Ultrasound Gel - MediChoice (M500812) due to bacterial contamination with
Burkholderia stabilis.
Products affected
Healthcare facilities are instructed to identify
affected lots and
immediately destroy products from affected lots. Affected product lot numbers distributed in Canada are below:
|
Product
|
Affected Lot Numbers
(destroy)
|
Initial Distribution Date
|
|
EcoGel 200 Ultrasound Gel – MediChoice (M500812) |
B029
|
March 26, 2021
|
|
B030
|
March 30, 2021
|
|
B031
|
April 5, 2021
|
|
B032
|
April 7, 2021
|
|
B040
|
April 26, 2021
|
|
B041
|
April 26, 2021
|
|
B048
|
May 7, 2021
|
|
B055
|
May 26, 2021
|
Additionally, healthcare facilities are instructed
to immediately stop use and quarantine all lots of the following ultrasound gels distributed under these brand names in these countries (and may be present in Canada):
-
EcoGel 200, by Eco-Med (Canada)
-
EcoGel 200, by Active Crystal Inc (Canada)
-
EcoGel 200, by Cardinal Health Canada
-
EcoGel 200, by CMCC Supply Centre and Bookstore (Canada)
-
EcoGel 200 Blue, by Christie Innomed Inc. (Canada)
-
EcoGel 200, by Dectro International (Canada)
-
EcoGel 200, by Medline Canada
-
EcoGel 200, by Dunbar Medical (Canada)
-
EcoGel 200, by Eastern Medical Supply Ltd (Canada)
-
EcoGel 200, by Henry Schein Arcona (Canada)
-
EcoGel 200, by Maranda Lauzon Inc (Canada)
-
EcoGel 200 Blue, by MJM Distributing (Canada)
-
EcoGel 200, by Ortho Canada (Canada)
-
EcoGel 200 and EcoGel 200 Blue, by STAT Healthcare Corporation (Canada)
-
EcoGel 200 and EcoGel 200 Blue, by The Stevens Company Limited (Canada)
-
EcoGel 200, by Strathroy Middlesex General Hospital (Canada)
-
EcoGel 200, by Trimedic Supply Network Ltd. (Canada)
-
EcoGel 200, by Vitality Depot (Canada)
-
Medico, by Medico Inc. (Canada)
-
Red Medical Ultrasound Gel Blue, by Red Medical Supplies Ltd. (Canada)
-
MediChoice Ultrasound Gel, by Owens & Minor (United States)
-
MediChoice Ultrasound Gel, by Mac Medical Supply Co Inc. (United States)
-
EcoGel 200, by Agencia Matamoros (Honduras)
-
EcoGel 200, by Global Medical Solution (NZ) LTD (New Zealand)
-
EcoGel 200, by Good-link Electronics Ltd. (Hong Kong)
-
EcoGel 200, by Medi’s Part Ltd. (Thailand)
-
EcoGel 200, by So medico SDN BHD (Malaysia)
-
Chattanooga Conductor USA, by DJO Global (United States)
-
Conductor Australia, by DJO Australia
-
DJO Conductor, by DJO Global (United States)
-
Liquasonic, by Athena Medical Products Inc. (United States)
-
Medline, by Medline Industries Inc. (United States)
-
NDC Eco-Gel 200 Ultrasound Gel, by NDC Inc. (United States)
-
Omni, by Accelerated Care Plus Leasing Inc. (United States)
-
Smart 200, by Smart Technology and Product Co., Ltd (Thailand)
Background Information
EcoGel 200 Ultrasound Gel - MediChoice (M500812) is a
non-sterile, medium viscosity, blue or clear ultrasound gel which is not indicated for sterile ultrasound procedures.
The manufacturer, Eco-Med Pharmaceuticals, Inc. initiated
a voluntary recall in Canada and globally of certain lots due to bacterial contamination. The source(s) of the bacterial contamination is currently unknown.
The effects of the bacteria,
Burkholderia stabilis, a member of the
Burkholderia cepacia complex (Bcc), vary widely, ranging from no symptoms at all to serious infections. Bcc bloodstream infections may result in sepsis and in certain cases, potentially death. The use of ultrasound gels manufactured by Eco-Med have been associated with at least 15 infections to date, as indicated in the manufacturer’s
voluntary recall letter, including bloodstream infections. Additional infections have been identified that may also be associated with the use of this ultrasound gel and investigation is ongoing. Potential routes of transmission leading to bloodstream infections associated with a contaminated non-sterile gel may have included the use of the gel for visualization prior to, in preparation for, or during an invasive procedure or application of the gel inside the sterile ultrasound probe sleeve during an invasive procedure using ultrasound guidance.