Health Products

VITAL SIGNS MONITOR 6000 SERIES (6300 BASIC) (2021-08-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
H. CONNEX SPOT MONITOR (7100 SERIES)71WE-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
I. CONNEX SPOT MONITOR (7300 SERIES)73CE-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
F. VITAL SIGNS MONITOR 6000 SERIES (CVSM 6700)67MCEP-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
G. VITAL SIGNS MONITOR 6000 SERIES (CVSM 6800)68MXEP-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
D. CONNEX INTEGRATED WALL SYSTEM (CIWS) (840000 STANDARD)84MXVEC-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
E. CONNEX INTEGRATED WALL SYSTEM (CIWS) (850000 WIRELESS)85MTVEC-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
K. CONNEX SPOT MONITOR (7500 SERIES)75CE-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
B. VITAL SIGNS MONITOR 6000 SERIES (6500 STANDARD)64MTPE-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
L. BRAUN THERMOSCAN PRO 600006000-200
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
C. VITAL SIGNS MONITOR 6000 SERIES (6500 WIRELESS)65MTXE-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
M. BRAUN PRO 600006000-900
More than 10 numbers, contact manufacturer. 
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
A. VITAL SIGNS MONITOR 6000 SERIES (6300 BASIC)63MXPE-B
More than 10 numbers, contact manufacturer.
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
J. CONNEX SPOT MONITOR (7400 SERIES)74CE-B

Introduction

Affected Products 
  1. VITAL SIGNS MONITOR 6000 SERIES (6300 BASIC)
  2. VITAL SIGNS MONITOR 6000 SERIES (6500 STANDARD)
  3. VITAL SIGNS MONITOR 6000 SERIES (6500 WIRELESS)
  4. CONNEX INTEGRATED WALL SYSTEM (CIWS) (840000 STANDARD)
  5. CONNEX INTEGRATED WALL SYSTEM (CIWS) (850000 WIRELESS)
  6. VITAL SIGNS MONITOR 6000 SERIES (CVSM 6700)
  7. VITAL SIGNS MONITOR 6000 SERIES (CVSM 6800)
  8. CONNEX SPOT MONITOR (7100 SERIES)
  9. CONNEX SPOT MONITOR (7300 SERIES)
  10. CONNEX SPOT MONITOR (7400 SERIES)
  11. CONNEX SPOT MONITOR (7500 SERIES)
  12. BRAUN THERMOSCAN PRO 6000
  13. BRAUN PRO 6000

Reason

Hillrom has become aware of customer complaint alleging that the speculum tip on the Braun PRO6000 was overheating. An investigation into the root cause of the speculum tip overheating determined that fluid ingress is causing the sensor to behave inconsistently, which ultimately does not allow the built-in safety mitigation to work correctly.