Health Products

Monaco (2021-08-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
(01)00858164002190(10) 5.40.00Monaco
Manufacturer
ELEKTA, INC.
1450 Beale street, suite 205
63303
MONACO

Introduction

Affected Products 

Monaco

Reason

Within offline Monaco, if a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams. When the plan is reloaded into planning, the user cannot re-assign the couch to the beams for calculation because the option is disabled. Therefore, the patient will be underdosed.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime