HEATER-COOLER UNIT HCU40 (2021-08-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturerHEATER-COOLER UNIT HCU40
Manufacturer
Maquet Cardiopulmonary GmbH
Kehler Str. 31
Rastatt
76437
GERMANY
70105.4054 (200-240V)

Introduction

Affected Products

HEATER-COOLER UNIT HCU40

Reason

Cleaning and/or descaling procedures for the internal water circuits can lead to crystalline deposits in the 3-way-valve promoting corrosion and malfunctions. In case of a malfunction of the 3-way valve, the valve does not close completely. In this case, a certain amount of warm water is not circulating in the main or patient circuit. Instead this warm water is guided back to the water tank leading to an increase of the water temperature and accelerated melting of the ice block. Consequently, the available time for the user to cool down the patient is reduced.