Health Products

ALINITY M SARS AMP KIT (2021-09-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not applicableALINITY M RESP-4 PLEX AMP KIT
Manufacturer
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES
Not applicableALINITY M SARS-COV-2 AMP KIT
Manufacturer
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES
09N78-090

Introduction

Affected Products

A. ALINITY M SARS-COV-2 AMP KIT
B. ALINITY M RESP-4 PLEX AMP KIT

Reason

Abbott Molecular inc. Has received reports of Alinity M SARS-CoV-2 EUA false positive results. Preliminary investigation has identified that potential carryover in the assay tray may contribute to false positive results. An evaluation of the current mixing protocol used during preparation of the PCR reaction mixture determined the current mixing parameters may result in potential overflow that could carry over into neighboring wells in the assay reagent tray. Because the assay parameters and the presence of the SARS-CoV-2 analyte are similar, Alinity M Resp-4-Plex customers are being included in this communication.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime