Health Products

Instrument Manager (2021-09-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Data Innovations LLC
120 Kimball Avenue Suite 100
South Burlington
05403
Vermont
UNITED STATES
Instrument ManagerV8.04 AND HIGHER

Introduction

Affected Products

Instrument Manager

Reason

After the upgrade of the data innovations advanced configurable delimited instrument driver to V8.04.0032, It was observed that the driver is not reading timestamps correctly when reading and parsing date/time values from a results file. When there are multifield date/time values for results, the driver is parsing both values but is discarding the time value resulting in a timestamp of 00:00:00 (midnight).