Health Products

CARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP (2021-09-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturerCARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
0998-00-0800-XX

Introduction

Affected Products

CARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP

Reason

1. In the presence of radiated RF disturbances, the Helium indicator on the cardiosave IABP would intermittently over-report the remaining helium capacity.
2. The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is inserted in the IABP, and the battery is physically removed while the battery is being charged.